Luldlum Model 12 Restoration

Recently I’ve been restoring a Ludlum Model 12 rate meter. What is that you might ask?

The best ratemeter on this side of Si Valley, that’s what!

I say this, because it’s one of those rare instruments built with “user serviceable” in mind, much like old tektronix oscilloscopes. Along with its bare, “don’t touch this lest you get shocked” circuit board, it contains trim potentiometers and for nearly everything one could wish to adjust; the high voltage power supply, the pulse discriminator, the integrator dividers, as well as a switch to change the integration speed and to reset the integrator entirely. But those are just big words for people who don’t know what the hell a survey meter is!

What is a rate meter / survey meter?

A survey meter is a radiation measurement device; something that provides for a measurement probe;

  • A high voltage power supply. Most radiation detectors (eg, geiger-muller tubes) require a high operating voltage, typically on the order of 700 to 2,500VDC.  Geiger-muller tubes typically require a potential of 400-700V, photomultiplier tubes a potential of 900 to 1500V, and proportional counters a potential of 1200 to 2500V.
  • A pulse discriminator. During a count events, an electrical pulse is formed across the radiation detector. On some detectors, such as neutron proportional counters, this magnitude of this pulse is co-relates to the detected particle’s energy. The discriminator is a circuit capacitvely coupled to the probe’s high voltage power supply which monitors these pulses, and triggers a count event if the pulse exceeds some magnitude. By adjusting the reference voltage, one can discriminate between say, a neutron, and a gamma ray.
  • An integrator. This is just some display mechanism that integrates the pulses, to give a “counts/minute” measurement. The accuracy of the integrator is of course, highly dependent on the length of time which it is active.
  • A speaker. Click-click. Click-click-click-click.

Naturally, being able to adjust all of these things means you can use a Ludlum-12 with any probe you wish!

My rate meter

Ludlum-12’s in working condition are notoriously hard to find, as they’re highly coveted instruments, and no one lets go of a working one easily! As such, I bought mine semi-broken and much beat-up. The meter I acquired had lost its voice at some point in its life, and was in much need of a face-lift.

WIN_20140906_103143

In all her pride and glory

WIN_20140906_103228 (2)

Analog dreams

By “losing its voice”, I mean that my meter had no ability to make sound. This is rather unfortunate, because while it might seem like only a minor inconvenience, those clicks could be the alarm one needs to learn whether or no one is in a radiation field!

The Ludlum-12 is an entirely analog instrument. Nowhere on its board exists any form of proprietary turing machine, which is rather fortunate as this makes it quite repairable. So where was the damage? Here!

1907898_10101016626678909_8599671716903897604_o

Ludlum-12 Audio Circuit, courtesy Andrew Seltzman

The audio circuit in this machine is rather simple. An astable multivibrator (CD4098) is triggered by the pulse discriminator (not labeled), which creates a few-kHz audio tone that’s fed out of Q2 into a NAND gate (CD4093). This NAND gate, being of the schmitt trigger type, squares-up the signal a bit, where it’s then further buffered by a second NAND gate and sent into a piezoelectric speaker. The audio “mute” switch is simply an SPDT toggle that holds the multivibrator’s RES2 (reset) pin low. These are the two ICs that somehow failed in this meter; likely due to some sort of electrostatic event given that they’re old and of fragile design. Interestingly, the multivibrator failed in such a way that pulses were able to make it out of Q2, but not periodic square waves. Weird. Replacing these ICs fixed the issue.

The Facelift

Whilst fixing this meter I was convinced by a fellow artist to make it snazzy.

WIN_20140908_175603

Get Funky

The above was done with rust-oleum enamel, masking tape, and a razor blade.

Testing

Though I will eventually use this as a scintillation meter, I needed a probe for testing purposes. Not having one on hand, I decided to use a Navy surplus GM-tube that had been collecting dust for some years. Since the Geiger tube runs at a high potential (700VDC, in this case) it’s required that the tube be insulated lest I shock myself while using it. My solution, given a lack of tools was a composite tube built from paper, epoxy, and vinyl tape. Given a small coat of polyurethane for strength, it’s just about as good as one made from aluminum!

WIN_20140906_103344

A DIY Composite Tube

WIN_20140906_103431

Fits my Geiger tube snugly

WIN_20140906_144321

And hosts a nice BNC connector

WIN_20140907_233542

To make one sexy probe

The Video

No sir I don’t want spicy Mexican; I want hot.

Add a dash of radium chloride please. ∎

← Previous post

Next post →

2 Comments

  1. Tim Burns

    Did you ever find an service manual with schematic for this model can’t seem to find one I also have an older 12 with all transistors circuit I need schmatic for, Ludlum only has the newer manuals.

    • ben hall

      Send Ludlum an e-mail with your serial number. They are *great* about responding and will send you the matching manual.

      thanks,
      ben

Leave a Reply to Tim Burns Cancel reply

Your email address will not be published.